Maintenance

Performance Assurance Checks

Performing

the Invasive

Pressure

Auto Check - The check for these modules is started every time the

Module Self-

system is switched ON and checks the complete module circuitry,

Test

including the sensitivity detection circuit.

 

Self-Test- The test signals lasts about 30 seconds and then the display

 

returns to the normal monitoring mode.

 

The M1006A Pressure self-test may be performed at any time as the

 

input switch switches the CAL divider to 0mmHg, and the pressure test

 

signals bypass the sensitivity detection circuit. From here, the signals

 

are processed in the same way as the patient signals.

 

The M1006B Pressure self-test may be performed at any time, as the

 

signal is output from a microprocessor based controller. From here, the

 

test signal is processed in the same way as a patient signal, testing the

 

complete signal path from the module to the display.

 

TEST - Press

 

 

 

 

 

 

 

Monitor Setup

 

 

 

Test Signals

 

 

RESULT - If the module is functioning correctly, the following systolic

 

and diastolic pressures are displayed without the alarms being activated:

Systolic 120 mmHg ADULT 60 mmHg PEDI/NEO

Diastolic 0 mmHg ADULT 0 mmHg PEDI/NEO

13-16 Maintenance

Page 382