Philips V26CT, V24CT manual Arrhythmia Alarms on the 78720 Arrhythmia Computer

Models: V26CT V24CT

1 392
Download 392 pages 63.56 Kb
Page 150
Image 150

Other Patients

Using Philips Patient Care System with an Arrhythmia Computer

Arrhythmia Alarms on the 78560A Arrhythmia Computer

Alarm

Minimum Condition Required for Alarm

 

 

RUN VPBs > X (**)

Run of more than X VPB's, where X is

 

adjustable from 2 to 9 using the Change

 

Limits display.

PACER NON-FUNCT

Missed beat without pace pulse in paced

(**)

patient

VPBs > XX/min (**)

More than XX VPBs per minute, where XX is

 

adjustable from 1 to 30 using the Change

 

Limit display.

 

 

Arrhythmia Alarms on the 78720 Arrhythmia Computer

Alarm

Minimum Condition Requiring an Alarm

 

 

EXTREME TACHY

Heart-rate >Tachy limit

 

 

RUN VPBs >9 (**)

Run of VPB's > 9 and heart-rate <esc< limit

 

set for VENT TACHY

 

 

RUN VPB's 3-9 (**)

Run of 3-9 VPB's

 

 

PACER NOT PACE (**)

Missed beat without pace pulse in paced

 

patient

 

 

VPB's > 23/min (**)

More than 23 VPB's/minute

 

 

VPB's > 11/min (**)

More than 11 VPB's/minute

 

 

VPB's > 5/min (**)

More than 5 VPB's/minute

 

 

4-12 Other Patients

Page 150
Image 150
Philips V26CT, V24CT manual Arrhythmia Alarms on the 78720 Arrhythmia Computer, Alarm Minimum Condition Requiring an Alarm