Maintenance

Performance Assurance Checks

Performing

Auto Check - The Blood Analysis Module performs a check when the

the Blood

system is switched on

Analysis

Self-Test- The Self-Test with the External Simulator M3634A may be

Module Self-

Test

performed at any time.

 

TEST - Insert the External Simulator into the cartridge slot of the

 

module.

 

RESULT - Either PASS or FAIL will be displayed on the Monitor to

 

indicate whether the module has passed the self-test or not. If the

 

module fails the self-test, try the procedure again. If it fails again, wait

 

half an hour and then try the procedure again. If the module continues

 

to fail the self-test the module should be replaced.

Performing

Auto Check - The recorder module performs a check when it is plugged

the

into the parameter module rack.

Recorder

Self-Test- The Recorder Module check may be performed at any time.

Module Self-

Test

TEST - Turn on the system power and plug in the recorder module

 

 

RESULT - The LED on the front panel indicates whether the module has

 

passed the self-test by flashing.

 

LED flashes twice - module has passed check LED flashes once - module

 

failed self-test. Take the module out and try the procedure again to

 

make sure there is a good connection. LED does not flash - module

 

processors are not working properly and the module should be

 

replaced

Performing

The data management database is tested during system start-up with an

the Data

SRAM checksum procedure. If there is an error found in the SRAM

Manage-

checksum, the system will restart (cold start) and all patient data will be

ment Data-

lost.

base Self-

 

Test

 

13-20 Maintenance