Adjusting Neonatal Event Review Settings

 

The Event Setup task window enables you to adjust the following items:

 

Trigger Mode

The trigger mode determines whether an

 

 

event is triggered by a physiological alarm or

 

 

by criteria defined specifically by the user.

 

Trig Threshold

The trigger threshold can be adjusted if the

 

 

user defined trigger mode setting has been

 

 

selected

 

Trigger Time

The trigger time can be adjusted if the user

 

 

defined trigger mode setting has been

 

 

selected.

Event

The event type (Apnea, Bradycardia, Desaturation) determines how a

Criteria

specific event is triggered and stored on the monitor.

Apnea Event

Apnea events are always triggered by a Red*** Apnea alarm. The Apnea

Settings

alarm is derived from the RESP parameter and occurs whenever

 

breathing ceases for a period exceeding the defined Apnea alarm limit

 

(usually 20 seconds).

 

The Apnea alarm limit and trigger time can be viewed but not adjusted in the Event Setup task window. These settings must be adjusted in Monitoring Mode, in the RESP task window.

Neonatal Event

Review

Neonatal Event Review 9-19