Philips V26CT, V24CT manual Monitor Installation Patient Safety

Models: V26CT V24CT

1 392
Download 392 pages 63.56 Kb
Page 338
Image 338

Monitor Installation

and Patient Safety

 

 

Controls and Connectors

Maximum Voltages

 

 

RS232 Connectors

±12 V

 

 

HDLC Connector

5 V

 

 

Video Out (9)

5 V

 

 

Flatscreen Remote Power Supply On/Off

15 V

 

 

Video Out (11)

60 V

 

 

Note—Do not apply a voltage of more than ±12V to any of these outputs.

The connectors on the rear panel of the computer module (together with their connecting function cards) are:

1.System Power Connector; on the M1046A Computer Module, this is a connector used to input the 60V dc power supply from the CRT display module. It is not for use on the M1046B Computer Module but is instead fitted with a stub connector.

2.Equipotential Grounding Terminal; this is used to connect the computer module to the hospital´s grounding system.

3.Philips monitoring network connectors (Philips Network Interface); these are Philips monitoring network connectors, used to input/output information to/from a Philips monitoring network.

4.Human Interface Connector (Utility CPU); this is a Philips HIL connector, used to input information from the keypad in the display module.

5.Nurse Paging Relay output; this is a mini-phone jack, used to connect to the hospital´s Nurse Paging System.

6.Analog Interface Connector; a 37-pin “D” type connector is used to output information to analog devices.

7.RS232 Connectors (RS232 Interface); these are 25-pin “D” type connectors, used to output information to printers or computers,

11-22 Monitor Installation and Patient Safety

Page 338
Image 338
Philips V26CT, V24CT manual Monitor Installation Patient Safety