Recording Functions

oxyCRG Recordings

oxyCRG Recordings

You can record the contents of the oxyCRG display on an M1116B Plug- In recorder.

Note—oxyCRG recordings can only be made on the M1116B Plug-In recorder. The M1116A Plug-In recorder, the M1117A four channel recorder, and the central recorder do not support oxyCRG recording.

You can manually request an oxyCRG recording, and you can configure an alarm recording of oxyCRG.

Manual request: An oxyCRG recording with a fixed delay of six minutes (corresponding to the display content) is produced.

Alarm recording: An automatic oxyCRG recording of six minutes of pre-alarm data and two minutes of post-alarm data is produced.

The oxyCRG recording is always 6 minutes prior to the most recent data on the screen. This enables you to have a hardcopy of events leading up to the current situation. When you stop recording, the recorder stops immediately, with the result that the most recent 6 minutes of patient data will not be recorded.

Note—OxyCRG alarm triggered recordings differ slightly from the oxyCRG episode recordings generated by Neonatal Events. All standard oxyCRG recordings are realtime recordings, i.e. when the alarm occurs. An oxyCRG episode recording is initiated by the user to document selected events retrospectively.