Philips V26CT, V24CT manual Care Operating Storage, Acms with Anesthetic Gas Module

Models: V26CT V24CT

1 392
Download 392 pages 63.56 Kb
Page 326
Image 326

Monitor Installation

and Patient Safety

 

 

Installation Information

Philips M1167/77A CMS for Critical, Anesthesia and Neonatal Care

 

 

Operating

Storage

 

 

 

 

 

 

 

 

 

Temperature Range

5 to 45°C (41 to 113°F)

-20 to 60°C (-4 to 140°F)

 

 

 

 

 

Humidity

up to 95% RH at 40°C (104°F)

up to 85% RH at 50°C (122°F)

 

 

 

 

 

Altitude Range

up to 4,600m (15,000 ft)

up to 4,600m (15,000 ft)

 

 

 

 

 

Philips M1165/66/75/76A CMS for Critical, Anesthesia and Neonatal

 

 

Care

 

 

 

 

 

 

 

Operating

Storage

 

 

 

 

 

 

 

 

 

Temperature Range

0 to 45°C (32 to 113°F)

-40 to 70°C (-40 to 158°F)

 

 

 

 

 

Humidity

up to 95% RH at 40°C (104°F)

up to 95% RH at 65°C (149°F)

 

 

 

 

 

Altitude Range

up to 4,600m (15,000 ft)

up to 15,300m (50,000 ft)

 

 

 

 

 

 

ACMS with Anesthetic Gas Module

 

 

 

 

 

 

Operating

Storage

 

 

 

 

 

 

 

 

 

Temperature Range

15 to 40°C (59 to 104°F)

-40 to 65°C (-40 to 149°F)

 

 

 

 

 

Humidity

up to 95% RH at 40°C (104°F)

up to 95% RH at 65°C (149°F)

 

 

 

 

 

Altitude Range

up to 3,048m (10,000 ft)

up to 15,300m (50,000 ft)

 

 

 

 

 

 

Philips M1205A V24, V26, V24C and V26C

 

 

 

 

 

 

Operating

Storage

 

 

 

 

 

 

 

 

 

Temperature Range

5 to 40°C (41 to 104°F)

-20 to 60ºC (4 to 140ºF)

 

 

 

 

 

Humidity

up to 95% RH at 40°C (104°F)

up to 90% RH at 60ºC (140ºF)

 

 

 

 

 

Altitude Range

up to 4,600m (15,000 ft)

up to 4,600m (15,000 ft)

 

 

 

 

11-10 Monitor Installation and Patient Safety