Philips V26CT, V24CT manual Recording Functions

Models: V26CT V24CT

1 392
Download 392 pages 63.56 Kb
Page 196
Image 196

Recording Functions

Realtime Wave Recordings

For central recorders, the requests are run and queued according to the priority for that recorder. See the operating guide of the recorder for information. For the M1116A/B Plug-In Recorder and the Philips M1117A Bedside Recorder, the recording prints when the recorder is free.

Note—If you are using a monitor connected to a Patient Information Center (78504 and 78508) and you are using an 78572 recorder, please observe the following rule:

Always select a valid waveform. Do not select “Blank” as the waveform to be recorded in the Select Recording Waves Task Window, as the recorder will not function. It will give the prompt “Central recorder not available”.

For the Monitor connected to the a Philips central station, realtime recordings must have at least one non-blank channel to generate a recording on the central recorder.

6-24 Recording Functions

Page 196
Image 196
Philips V26CT, V24CT manual Recording Functions