Combining Data

Combining Data

 

In some situations, the patient name and/or ID in the monitor and

 

module may not match (see Append New Data display on previous

 

page). In other cases, one or both names may be blank. The Data

 

Transfer Combine Task Window enables you to combine files when the

 

demographic data is missing or different.

Time

Data in a monitor can be stored at either 1-minute or 12-second

Conversion

resolution, depending on the monitor's configuration. Data in a module

 

is stored only at 1-minute resolution. Time conversions are done

 

automatically when the resolutions are different.

 

When converting data from 12-second to 1-minute resolution (that is,

 

from a monitor to a module), one data sample out of 5 is selected to

 

represent the minute. The sample selected is the one on the minute

 

boundary.

 

When transferring data from 1-minute to 12-second resolution (from a

 

module to a monitor), one valid sample and 4 blanks will be used for

 

each minute. The valid sample will be placed on the minute boundary.

Database

Transfers to the Module

Conversion

A module contains enough memory to store 24 hours of data for 16

 

parameters at 1 minute resolution. If the data to be transferred does not

 

fit in the module, then the number of hours will be reduced to fit the

 

available memory.

 

The following table shows the conversion between the monitor and

 

module databases during a transfer to the module.

Data Transfer

10-14 Data Transfer