Philips V24CT, V26CT manual Alarm Reminder Nurse Call Relay Signal Behavior

Models: V26CT V24CT

1 392
Download 392 pages 63.56 Kb
Page 171
Image 171

Alarm Setup

The Nurse Note—The nurse call relay is not available for the V24CT and V26CT.

Call Relay

The nurse call relay is a paging system which warns the nurse if an alarm occurs on a monitor. It becomes active in one of four situations depending on the way it has been configured:

Red only alarms

Red and Yellow alarms

Red and INOP alarms

Red and Yellow INOP alarms.

The nurse call relay is not active when the monitor’s alarm capability is switched off.

Note—The delay time to the nurse call relay output is max. 1 sec.

When an alarm has been silenced using the Silence/Reset key, and the

audible 2 or 3 minute alarm reminder is configured to be either On or Re- alarm, the nurse call relay signals will behave as follows:

Alarm

Reminder

 

Nurse Call Relay Signal Behavior

Reminder

Time

 

 

 

 

 

 

 

 

 

On

2 or 3 mins

The nurse call relay signals are repeated

 

 

for a duration of 6 seconds every 2 or 3

 

 

minutes unless the alarm condition is

 

 

eliminated

 

 

 

 

 

Re-alarm

2 or 3 mins

The nurse call relay signals are activated

 

 

after 2 or 3 minutes. The alarm tone and

 

 

relay signals run continuously until they

 

 

are interrupted by pressing the

 

 

 

Silence/Reset

key.

 

 

 

 

 

Alarm Functions 5-15

Alarm Functions

Page 171
Image 171