Philips V24CT, V26CT manual M1094B Display Module, Controls, M1092A, M1094A, Displays

Models: V26CT V24CT

1 392
Download 392 pages 63.56 Kb
Page 341
Image 341

Controls and Connectors

M1094B Display Module

 

Maximum Voltages

 

 

 

System Power Connector

 

60 V

 

 

 

Human Interface Connector

 

12 V

 

 

 

Local Power Connector

 

240 V

 

 

 

Video Out Connector

 

5 V

 

 

 

Video In Connector

 

5 V

 

 

 

Controls of the

1. Video Termination Switch; thus switch is used to terminate the

M1092A,

video signal at this display: this must be switched to the “75

M1094A and

position if only a master display is used and must be switched to the

M1094B

“HIGH IMP.” position if a second, duplicate display is used.