Before commencing monitoring on a patient, carry out the following checks on the system:

Check for any mechanical damage

Check all the external leads, plug-ins and accessories

Check all the functions of the instrument which will be needed to monitor the patient, and ensure that the instrument is in good working order.

Do not use the system for any monitoring procedure on a patient

if you identify features which demonstrate impaired functioning of the instrument. Contact the hospital biomedical engineer or the Philips Service Engineer.

Maintenance

Frequency

Source of Information

 

 

 

 

 

 

Full performance checks

At least every two years, after

See “Testing and Maintaining the

including safety checks.

any repairs, or as needed.

CMS” in the “CMS Service Guide”

 

 

or “Testing and Maintaining the

 

 

V24 and V26“in the “V24 and V26

 

 

Service Guide”.

 

 

 

Synchronization of the monitor

If defibrillator is used or

See the Philips M1722/3/4A

and defibrillator.

configured for intended use: at

Codemaster XL/XL+/XE

 

least once a year, after any

Defibrillator Service Guide.

 

repairs, or as needed.

 

 

 

 

Preventive maintenance

Perform the NBP and Sidestream

See “Testing and Maintaining the

 

CO 2 Module Performance test

CMS” in the “CMS Service Guide”

 

blocks every year.

or “Testing and Maintaining the

 

Perform the Preventive

V24 and V26“in the “V24 and V26

 

Maintenance Procedure for the

Service Guide”.

 

M1180A #A01/A02 Wall Mount

 

 

Assembly every year.

 

 

 

 

11-14 Monitor Installation and Patient Safety

Page 330
Image 330
Philips V26CT, V24CT manual Maintenance Checks, Maintenance Frequency Source of Information