The Philips Patient Care System is a network which transports information. If your system is connected to such a monitoring network you can send information to other systems or central stations, and receive information from them.

One aspect of this information transmission capability is called Other

Patients.

The Other Patients setup, based on nursing care assignments, allows you to view data from another bed in your group or the “ALL” group.

If your monitoring system is connected to the monitoring network, you can have 7 waves transmitted to the central station*. Only the primary wave is displayed on the main screen at the central station. Each Other Patients screen of any device connected to the monitoring network starts with the primary wave.

If you have ECG turned on, it is always selected as the primary wave. If it is not turned on you can select a primary wave in the System Waves Task Window under Monitor Setup.

*If EASI 12-Lead ECG is configured ON only 5 waves can be transmitted to the central station.

The Other

The softkeys in the Other Patients Selection Window contain the bed

Patients

labels of the beds currently available for viewing on your monitor.

Selection

Pressing a softkey leads you into the Other Bed Task Window for that

Window

particular bed. This task window contains detailed patient information

 

including alarms and INOPs.

 

If your bed is configured to receive alarms and the beds in your Other

 

Patients group are configured to transmit alarms, the softkeys in the

 

Other Patients Selection Window contain alarm information under the

 

bed label. Alarm severity is indicated in the second line of the softkey

 

label by ** for a yellow alarm and by *** for a red alarm. Additionally, the

 

softkey will have a red or a yellow frame if a bed is in alarm.

Other Patients

Other Patients 4-3

Page 141