Installation Information

Environ-

To ensure a completely safe electrical installation, follow the instructions

ment

described later in this section. The environment where the system will be

 

used should be reasonably free from vibration, dust, corrosive or

 

explosive gases, extremes of temperature, humidity, and so on.

 

For a cabinet mounted installation, allow sufficient room at the front for

 

operation and sufficient room at the rear for servicing with the cabinet

 

access door open.

 

Approximately 15 minutes after switch-on, the patient monitor and the

 

Anesthetic Gas Module operate within specifications at ambient

 

temperatures shown in the tables below. These specifications do not

 

apply to the plug-in modules; they are specified for different ambient

 

temperatures.

 

Ambient temperatures that exceed these limits could affect the accuracy

 

of these instruments and cause damage to the components and circuits.

 

Allow at least 2 inches (5cm) clearance around the instruments for proper

 

air circulation.

 

Refer to the tables below for a summary of the environmental

 

specifications of the different system combinations:

Monitor Installation

and Patient Safety

 

 

Monitor Installation and Patient Safety 11-9