Philips V26CT, V24CT manual AC and DC Battery Operation

Models: V26CT V24CT

1 392
Download 392 pages 63.56 Kb
Page 356
Image 356

AC and DC (Battery) Operation

Information

and V26CT

Battery

(V24CT

 

 

AC and DC (Battery) Operation

You should only use Philips 40488A 12 Volt sealed lead-acid batteries obtained from Philips Medical Systems for use with the Philips M1205A V24CT. Use of other camcorder-type batteries could result in inaccurate estimates of battery capacity remaining or in shorter battery life.

Warning

When connecting a non-medical device such as a printer to the RS-232 Interface of V24CT, the monitor must be properly grounded, or the non-medical device properly isolated. This must be done in order to prevent the possibility of an electrical shock hazard (to the user and or patient) from excessive ground leakage currents from the non-medical device such as a printer, and to maintain compliance with IEC 601-1-1 where applicable. This hazard is especially prevalent when the V24CT is operated in battery mode.

In normal AC line operation grounding the monitor is accomplished by ensuring that the AC line cord is connected from the monitor to a properly grounded AC receptacle. Alternatively and additionally, proper grounding may be achieved by attaching a safety grounded wire to the threaded safety ground stud found on the rear of the monitor.

Isolation may be achieved by powering the printer through a safety isolation transformer.

12-2 Battery Information (V24CT and V26CT only)

Page 356
Image 356
Philips V26CT, V24CT manual AC and DC Battery Operation