Installation Information

Installation Information

Power

The CMS systems and the Anesthetic Gas Module can be operated from

Source

an AC source of 100 - 120V ± 10%, or 220 - 240V ± 10%,

Require-

50 - 60Hz.

 

ments

Typical input currents for the monochrome and color systems are:

 

 

Monochrome CRT

100 - 120V = 2.2 amps

 

 

200 - 240V = 1.2 amps

 

Color CRT

100 - 120V = 2.6 amps

 

 

200 - 240V = 1.4 amps

 

Color Flatscreen

110 - 120V = 2.0 amps

 

 

200 - 240V = 1.0 amps

Typical input current for the Anesthetic Gas Module is:

100 - 240V = 1.0 amps

Typical input current for the V24 and V26 is:

90 - 250V = 150VA

Grounding To protect the patient and hospital personnel, the cabinet of the installed

the System equipment has to be grounded. The equipment is supplied with a detachable 3-wire cable which grounds the instrument to the power line ground (protective earth) when plugged into an appropriate 3-wire receptacle. If a 3-wire receptacle is not available, consult the hospital electrician.

Warning

Do not use a 3-wire to 2-wire adapter.

Monitor Installation

and Patient Safety

 

 

Monitor Installation and Patient Safety 11-5