Changing the Patient Category

NBP Safety Values for Differing Patient Categories

Safety Value

 

Patient Category

 

Adult

 

Pedi

 

Neo

 

 

 

 

 

 

 

 

 

Max. deflation time

80s

 

80s

 

60s

 

 

 

 

 

 

Max. measurement time

100s

 

100s

 

60s

 

 

 

 

 

 

Overpressure protection1

300 mmHg

 

300 mmHg

 

150 mmHg

 

 

 

 

 

 

Cuff Size

A to E

 

Any Cuff

 

1 to 4 & A

 

 

 

 

 

 

Default Inflation Pressure

165 mmHg

 

125 mmHg

 

100 mmHg

 

 

 

 

 

 

1The figure given for overpressure protection is the maximum pressure the cuff can reach for 2 seconds before it deflates.

NBP Recommendations

Philips Medical Systems makes the following recommendations when selecting a patient category for NBP:

1. Choose the category with the most appropriate safety values for your patient.

2. Reduce the stress on the patient by choosing the appropriate cuff size (a cuff that is too large prolongs inflation time).

3. Use a repetition time appropriate to the patient.

ECG

The following factors can cause the cardiotach to calculate an incorrect

 

heart rate:

 

• Morphology of the ECG,

 

Weak signal,

 

Paced patients.

Morphology of

Philips offers two different ECG algorithms to allow for differing

the ECG

morphologies.