Neonatal Event

Review

 

 

Viewing Neonatal Events

If the event storage is full, the newest (49th1) event will enter the graph on the right hand side and push the oldest (1st) event off the graph on the left hand side. This information cannot be retrieved. To document the event review before the event storage becomes overfull, you should initiate an event summary recording.

To free up space in the event storage and remove unwanted events from the Event Review graph, you can delete events in the oxyCRG Episode task window.

Operating In the Event Review task window you can:

Controls

1.Select the previous event.

2.Select the next event.

3.View an oxyCRG Episode for the selected event.

4.Record the event summary.

5.Print the event summary

6.Enter the Event Setup Window to change the neonatal event trigger conditions.

Selecting an

Selecting the

Previous Event

and/or

Next Event

softkeys

Event

moves the cursor from one event to the next in both directions.

 

Alternatively, the arrow keys on the control panel can also be used.

Once the cursor has been positioned on an event, the date and event time is shown and the event values specific to that event are displayed in the three boxes located on the right hand side of the task window.

Viewing an

Selecting the

oxyCRG Episode

softkey enters the oxyCRG Episode

oxyCRG

task window, where you can review the oxyCRG episode for the event

Episode for

currently selected. At this point, and not before, an event can be deleted

the Selected

permanently from the event storage.

Event

 

 

 

 

 

 

 

 

 

 

 

 

 

1. 25th for V24 and V26 Patient Monitors