Philips V24CT, V26CT manual Battery Care and Maintenance

Models: V26CT V24CT

1 392
Download 392 pages 63.56 Kb
Page 363
Image 363

Battery Care and Maintenance

Battery Care and Maintenance

 

The Philips 40488A sealed lead-acid batteries used in the V24CT provide

 

optimum performance and battery life when properly cared for. Proper

 

handling of batteries is essential to the life of the battery and safety of the

 

user.

Storage

• Store batteries in a cool dry area. A refrigerator is an excellent place

 

to store batteries. Allow batteries to return to room temperature

 

before charging or use.

 

• The charge in the batteries gradually deteriorates even during

 

storage. They will self-discharge 4-5% for each month in storage at

 

25°C.

 

• Batteries should be stored in a fully charged state.

 

• A stored battery loses capacity over time. To minimize loss of

 

capacity recharge battery at least once every 6 months. Batteries

 

can be charged in the monitor or by using a battery charger.

 

Note—Recharge the batteries after using them. It is especially important

 

to avoid allowing the batteries to stay in a fully discharged state for long

 

periods of time. Simply plug the monitor into AC line power after

 

transporting a patient or when the monitor is not in use. If possible, avoid

 

discharging the battery completely.

Replace batteries with known charged batteries in the following conditions:

Batteries discharge quickly when being used

Batteries’ runtime after first “RECHARGE BATTERIES” alarm is less than ten minutes

Batteries do not charge to full capacity (indicated by the monitor's battery fuel gauge)

Information

and V26CT

Battery

(V24CT

 

 

Battery Information (V24CT and V26CT only) 12-9

Page 363
Image 363
Philips V24CT, V26CT manual Battery Care and Maintenance