Monitor Installation

and Patient Safety

 

 

Maintenance Checks

 

Patient

Inspect the patient cables and leads and their strain reliefs for general

Cables and

condition. Examine cables carefully to detect breaks in the insulation

Leads

and to ensure that they are gripped securely in the connectors of each

 

end to prevent rotation or other strain.

 

To ensure operation of the Anesthetic Gas Module within specified

 

limits:

 

• calibrate the Anesthetic Gas Module at least once every 12

 

months, or whenever the validity of the readings is in doubt.

 

• check the ventilator fan in the Anesthetic Gas Module for proper

 

operation every 6 months.

 

All checks which require the instrument to be opened must be made by

 

qualified hospital service personnel. Safety and maintenance checks can

 

also be made by Philips personnel.

 

 

 

 

 

Warning

 

 

 

Failure on the part of the responsible individual hospital or

 

institution employing the use of this equipment to implement a

 

satisfactory maintenance schedule may cause undue equipment

 

failure and possible health hazards.

 

 

 

 