Alarm Display

Latching Alarms - The alarm continues when the alarm condition is eliminated. Latching alarms can be reset by pressing the

Silence/Reset key. Alternatively, all latching alarms can be turned off using the Suspend Alarms softkey or the Suspend hardkey. Latching alarms for individual parameters can be turned off in the Adjust Alarms Task Window or the Alarm Limits Task Window.

Non-latching Alarms - These automatically reset when the alarm condition is eliminated. INOPs are always non-latching.

Alarm Reminder - This only functions with alarms and not INOPs. (It does not work for arrhythmia alarms.) It affects the alarm sound, lamp, nurse call relay and the recorder, but not the alarm messages. The alarm reminder and reminder time can be configured as follows:

Alarm Reminder

Reminder Time

 

 

Off

Not Applicable

 

 

On

2 or 3 mins

 

 

Re-alarm

2 or 3 mins

 

 

If you have acknowledged an alarm by pressing the Silence/Reset key and the alarm condition continues, then the system will behave according to the following configurations:

Alarm

Reminder

Alarm Behavior

Reminder

Time

 

 

 

 

Off

----

No alarm reminder.

 

 

 

On

2 or 3 mins

Alarm reminder tone

 

 

Alarm lamp

 

 

No new alarm recording

 

 

Nurse call relay signals

 

 

The alarm tone is reannounced for a

 

 

duration of 6 seconds every 2 or 3

 

 

minutes unless the alarm condition is

 

 

eliminated.

 

 

 

 

 