Image 385

Performing

the CO2

Module Self-

Test

Performing the Temperature Module Self- Test

Performance Assurance Checks

Auto Check - None

Self-Test- The self-test may be performed at any time. The test signals last about 30 seconds, and then the display returns to the normal monitoring mode.

TEST - If a transducer is connected to the module, the following conditions must be met:

The motor in the transducer is running.

No heater malfunction is reported.

Press

Monitor Setup

 

 

 

Test Signals

 

 

RESULT - If the module is functioning correctly the values will be displayed without the alarms being activated and should read as follows:

ET CO2 40 mmHg (6.0kPa)

IMCO2

0 mmHg (0.0kPa)

Maintenance

AWRR

25 rpm

 

Auto Check - None

 

Self-Test- The Temperature module self-test may be performed at any time as the switch network switches to the reference resistors for the duration of the check. The system connects the reference resistors to the beginning of the processing circuitry, and processes this test value in the same way as the patient values are processed. The test signals last about 30 seconds and then the display returns to the normal monitoring mode.

TEST - Press

Monitor Setup

 

 

 

Test Signals

 

 

RESULT - If the module is functioning correctly, the following temperature will be displayed without the alarms being activated:

Temp 40 C (+0.1C)

Maintenance 13-19