Using Philips Patient Care System with an Arrhythmia Computer

Arrhythmia Alarms on the 3150A/3153A Arrhythmia Computer

Message

Minimum Condition Requiring an Alarm

 

 

*** ASYSTOLE

No QRS for 4 consecutive seconds

 

 

*** V-FIB/TACH

Fibrillatory wave for 4 consecutive seconds

 

 

*** V-TACH

Consecutive PVCs >/= V-Tach Run limit and

 

HR > V-Tach HR limit

 

 

*** TACHY yyy > xxx

Heart Rate (yyy) greater than the Extreme

 

Tachy limit (xxx)

 

 

*** BRADY yyy < xxx

Heart Rate (yyy) less than the Extreme

 

Brady limit (xxx)

 

 

** NON-SUSTAIN VT

A run of Vs having a ventricular HR>V-Tach

 

HR limit, but lasting for less than the V-Tach

 

Run limit

 

 

** VENT RHYTHM

A dominant rhythm of adjacent Vs > vent

 

rhythm limit and ventricular HR<V-Tach HR

 

limit

 

 

** RUN PVCs

Run of PVCs greater than 2

 

 

** PAIR PVCs

Two consecutive PVCs between non-PVCs

 

 

** PACER NOT CAPT

No QRS for 1.75 x the average R-R interval

 

with Pace Pulse (paced patient only)

 

 

** PACER NOT PACE

No QRS and Pace Pulse for 1.75 x the

 

average R-R interval (paced patient only)

 

 

** PAUSE

No beat detected for 1.75 x average R-R

 

interval for HR <120, or no beat for 1 second

 

with HR >120

 

 

** SVT

Run of SVPBs >/= SVT Run limit and with

 

SVT Heart Rate greater than the SVT HR

 

limit

 

 