Philips V26CT, V24CT manual Overview, System Self-test Values Module Test Numeric Test Waveform

Models: V26CT V24CT

1 392
Download 392 pages 63.56 Kb
Page 380
Image 380

Maintenance

Performance Assurance Checks

Procedure

1. Choose a patient category (Adult/Pedi/Neo) by pressing:

Overview

 

Monitor Setup

 

 

 

 

Config Sets

 

 

 

 

Change Pat Ca

teg

 

 

 

 

 

 

 

 

2. Display 30 seconds of the waveforms and numerics of all

 

connected modules by pressing:

 

 

 

 

 

 

Monitor Setup

 

 

Test Signals

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3. Repeat Step 2 as necessary to check the individual test waveforms

 

and numerics for the category.

 

 

 

 

 

4. Use Step 1 to choose a new patient category and repeat Steps 2

 

and 3.

 

 

 

 

 

 

 

 

 

 

See the following table for test values.

 

 

 

 

 

System Self-test Values

 

 

 

 

 

 

 

 

 

 

 

 

 

Module

Test Numeric

 

Test Waveform

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ECG and ECG/

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

RESP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ECG

100 bpm in ADULT mode

 

Simulated Square Wave

 

125bpm in NEO/PED mode

 

 

 

 

 

 

 

 

 

 

 

 

 

RESP

15 rpm in ADULT mode

 

Simulated Wave

 

30 nrpm in PEDI mode

 

 

 

 

 

 

 

 

 

 

55 rpm in NEO mode

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pressure

 

 

 

 

 

 

 

Simulated square wave

 

 

 

 

 

 

 

 

 

 

 

Systolic

120 mmHg ADULT

 

 

 

 

 

 

 

 

 

 

60 mmHg PEDI/NEO

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Diastolic

0 mmHg ADULT

 

 

 

 

 

 

 

 

 

 

0 mmHg PEDI/NEO

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SpO2/Pleth

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SpO2

100%

 

 

 

 

 

 

No Wave

 

 

 

 

 

 

 

 

Pleth

Wave on display

 

Simulated wave

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

13-14 Maintenance

Page 380
Image 380
Philips V26CT, V24CT manual Overview, System Self-test Values Module Test Numeric Test Waveform