Types of Transfer

Transferring data from a source monitor via a transit monitor to a destination monitor requires special handling if:

The source monitor and the destination monitor are:

CMS Release G or later or

V24 or V26 Release A.0 or later and

The transit monitor is:

CMS Release E or F

V24 Release E

CTS

1.Plug the Data Transfer Module into the rack of the source monitor.

2.Press the Module hardkey on the module and transfer all data to the module.

3.Plug the Data Transfer Module into the rack of the transit monitor.

4. Select Collect New Data in the Data to Module task window to collect new data to the module without clearing the module.

5.Remove the module from the transit monitor and plug it into the destination monitor.

6. Press the Monitor hardkey on the module and transfer the data to the monitor.

Note—With specific configurations, data loss can occur when transferring data to a V24 or V26 Release B.0 via the Data Transfer module.