Philips V26CT, V24CT manual M1046B, ECG Output and Defibrillator Marker Input

Models: V26CT V24CT

1 392
Download 392 pages 63.56 Kb
Page 56
Image 56

The CMS and V24 and V26 Patient Monitors

The CMS Computer Modules

M1046B

The M1046B Computer Module is for use with the Philips M1167/77A

Computer

CMS.

Module

 

Maximum Voltage

ECG Output and Defibrillator Marker Input

ECG Output and Defib. Marker Input

3.2 V p-p

 

 

The parameter modules are inserted into one or several satellite racks. They cannot be attached to the front of the computer module.

Applies to both the M1046A and M1046B Computer Modules.

The ECG output and defibrillator marker input connector on the computer module is used during cardioversion to provide an ECG waveform for synchronizing a defibrillator. It is also used to receive a marker pulse from the defibrillator for indicating the defibrillator discharge in the ECG waveform on the monitor display.

If you are using a Philips defibrillator without a built-in recorder, the marker pulse from the defibrillator is returned to the monitor. If the Philips defibrillator has a built-in recorder, then the marker pulse is returned to that recorder rather than to the monitor. The ECG output can also be used for other applications, such as synchronizing intra- aortic balloon pump systems.

1-34 The CMS and V24 and V26 Patient Monitors

Page 56
Image 56
Philips V26CT, V24CT manual M1046B, ECG Output and Defibrillator Marker Input