Philips V26CT, V24CT manual Setting up your Monitor

Models: V26CT V24CT

1 392
Download 392 pages 63.56 Kb
Page 122
Image 122

Setting up your Monitor

Changing the Patient Category

These safety values effect the measurement range of each category. The Adult category has the greatest range (systolic 30 to 270 mmHg) and the Neonate has the smallest (systolic 30 to 130 mmHg). Therefore you

cannot use Neonate or Pediatric if the expected pressure values are outside the measurement range for that category.

The following tables list the measurement ranges and safety values for each patient category.

NBP Measurement Range (in mmHg) for Differing Patient

Categories

Measurement Range

 

 

Patient Category

 

 

Adult

 

Pedi

 

Neo

 

 

 

 

 

 

 

 

 

 

 

 

 

Systolic

30 to 270

 

30 to 180

 

30 to 130

 

 

 

 

 

 

 

 

 

Mean

20 to 255

 

20 to 160

 

20 to 120

 

 

 

 

 

 

 

 

 

Diastolic

10 to 245

 

10 to 150

 

10 to 100

 

 

 

 

 

 

 

 

 

Cuff Size

A to E

 

Any Cuff

 

1 to 4 & A

 

 

 

 

 

 

 

 

 

Default Inflation Pressure

165 mmHg

 

125 mmHg

 

100 mmHg

 

 

 

 

 

 

 

 

NBP Safety Values for Differing Patient Categories

 

 

 

 

 

 

 

 

 

 

Safety Value

 

 

 

Patient Category

 

 

 

Adult

 

 

Pedi

 

Neo

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Continuous Pressure

 

15 mmHg

 

 

5 mmHg

 

5 mmHg (for

 

 

 

(for 120s)

 

 

(for 120s)

 

60s)

 

 

 

 

 

 

 

 

 

 

Time to inflate

 

40s

 

 

20s

 

20s

 

from 0 to 60 mm Hg

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Max. inflation time

 

60s

 

 

40s

 

30s

 

 

 

 

 

 

 

 

 