Page 152
Image 152

Other Patients

Using Philips Patient Care System with an Arrhythmia Computer

Message

Minimum Condition Requiring an Alarm

 

 

** R-ON-T PVCs

For HR <100, a PVC with R-R interval <1/3

 

the average interval followed by a

 

compensatory pause of 1.25 x average R-R

 

interval or 2 such Vs without a

 

compensatory pause occurring within 5 min.

 

of each other. (When HR >100, 1/3 R-R

 

interval is too short for detection.)

 

 

** VENT BIGEMINY

A dominant rhythm of N, V, N, V

 

(N=supraventricular beat, V=ventricular

 

beat)

 

 

** VENT TRIGEMINY

A dominant rhythm of N, N, V, N, N, V

 

(N=supraventricular beat, V=ventricular

 

beat)

 

 

** PVCs > xxx/MIN

PVCs within one minute exceeded the PVCs

 

/min limit (xxx)

 

 

** MULTIFORM PVCs

The occurrence of two differently shaped

 

Vs, each occurring at least twice within the

 

last 300 beats as well as each occurring at

 

least once within the last 60 beats

 

 

** HR yyy > xxx

Heart Rate (yyy) greater than the upper HR

 

limit (xxx)

 

 

** HR yyy < xxx

Heart Rate (yyy) lower than the lower HR

 

limit (xxx)

 

 

** IRREGULAR HR

Consistently irregular rhythm (irregular R-R

 

intervals)

 

 

4-14 Other Patients

Page 152
Image 152