Getting Started

Attaching the Patient

 

 

 

 

 

 

 

 

 

 

 

Power

If a power failure occurs, any alarm conditions that were stored in the

Failure

monitor will NOT be retained when power is restored. Settings such as

 

alarm limits stored in “T” modules ARE retained - see “Parameter

 

Settings Transfer” on page 3-61.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

C aution

 

 

 

 

 

 

 

You are advised to check all the monitor settings after any power failure.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The monitor reacts according to the length of the power failure:

 

Less than 60 seconds

Patient Data Management data and all

 

 

 

 

 

 

 

settings retained.

 

Less than 3 hours

Patient Data Management data retained. All

 

 

 

 

 

 

 

settings revert to user defaults.

 

More than 3 hours

Patient Data Management data not retained.

 

 

 

 

 

 

 

All settings revert to user defaults. Time

 

 

 

 

 

 

 

setting may be lost if not networked.

Patient

If the monitor is connected to a Patient Information Center (78504 and

Information

78508) and you select the same lead for ECG channels 1 and 2, the wave

Center

is not displayed in channel 2 of the Patient Information Center.

Monitor

If you want to stop patient monitoring, but keep all settings and Patient

Standby

Data Management information, you can switch the monitor into

 

“Monitor Standby”:

 

 

 

 

 

 

1.

Press

 

or

 

 

 

Alarms / Volume

Alarms

 

2.

Press

 

 

 

.

 

 

Monitor Standby

 

 

 

 

 

 

 

 

 

 

 

 

 

All waves and numerics disappear from the display, and all monitoring is suspended. The word “Standby” appears in big letters on the screen.

2-10 Getting Started