Recording Functions

Procedure Recordings

Procedures

WedgeEdit Wedge

ST Recordings

See the Pressure Section for details on Pulmonary Artery Wedge Pressure (PAWP) Measurements.

When a PAWP measurement has been performed, the PAWP waveform can be recorded by pressing the Record Curve softkey. If displayed, the CO2 or RESP waveform is also recorded.

The recording strip is also annotated with the date, time, and wedge value.

Recordings of ST waves and their reference waves can be made on the bedside or central recorders. They are initiated from the ST Analysis Task Window. An ST recording shows the ST beat(s) from the cursor spot in the upper channel, and the corresponding reference beat(s) in the lower channel, with the measurement values, net change, and measurement points listed in the annotation.

ST values transferred by the Philips M1235A Data Transfer Module are indicated by a “T” after the time. Because the ST waveforms are not transferred, no waveforms will be recorded.

6-18 Recording Functions