Configuring

the Other

Patients

Controls

Philips Patient Care System

With the Other Patients Controls you can place your own bed into one of eight groups: A, B, C, D, E, F, G or the ALL group, based on nursing care assignments within a unit.

The Other Patients Controls Task Window allows you to change the Send/ Receive alarms configuration for your bed.

Note—When a bed is assigned to the ALL group, alarms can be received from and sent to beds in group A-G as well as the other beds in the ALL group. When a bed is assigned to any group in A-G, alarms are received from and sent to any bed in the same group and also to beds in the ALL group.

Note—The bed will automatically be assigned to the ALL group if the connection to the Philips monitoring network is interrupted (for example, if the Philips monitoring network cable at the back of the monitor is disconnected, or the central station is switched off) and the monitor comes back on a different branch.

Note—Any changes made to Other Patients settings will be stored for up to three hours after the monitor is switched off.

Other Patients

Other Patients 4-7

Page 145
Image 145