Monitor Maintenance

Maintenance

Perform a

System Self-

Test

The following lists how to perform the system self-test.

1.Examine all controls for physical condition. During the course of this examination, ensure that the controls perform their proper function

2.Check that the parameter module LED's light up when first plugged in.

3.Confirm that the display lights up and displays a Main Screen.

4.Press the ECG key on the front of the ECG module in a system that is turned On. This brings up the ECG Task Window and verifies that the System board and integral rack are communicating with the plug-in module. If the ECG module does not respond, perform the ECG module Functional self-test.

5.Ensure correct operation (visually and audibly) of all alarms using, if required, the User Guide and a patient simulator.

6.If possible connect the ECG parameter module of the unit to an ECG simulator and verify that an adequate trace is received at each patient lead selection.

7.Press Monitor Setup , Monitor Revision , Show H W Rev . In this screen, press Show H W Rev . This provides a realtime list of the modules in the rack. Remove a module and observe the list changes to delete the module you removed. Re-insert the module and observe the list changes to include the module.