Symbols to Indicate Key Functions

Data Transfer Module

When the module is unplugged, it retains data for at least one hour.

After data is transferred to the monitor, the module is automatically erased. This is to ensure that the module is empty when it is used for the next transfer.

As detailed in the table below, some versions of this module now use symbols, instead of words, to indicate the function of some keys. If the monitor’s Reference Manual tells you, for example, to select the Monitor key, you should press the key marked with the corresponding “monitor” symbol.

The design change also means that you will now find the module’s product number on the rear of the module’s housing, not on the front.

Symbol

Name

Function

Which Modules?

 

 

 

 

 

Monitor

initiates transfer from module to

Data Transfer Module (M1235A)

 

 

monitor

 

 

 

 

 

 

Module

initiates transfer from monitor to

 

 

 

module

 

 

 

 

 

What is Transferred

The module is capable of storing and transferring up to 24 hours of trend data and demographic information. When a transfer is initiated, the following data is transferred:

Continuously monitored parameters (for example, ECG)

Resampled Values

Aperiodic measurements (for example, CO)

All patient demographic information except bed number

The time at which calculations were performed in the Hemodynamic, Oxygenation, and Ventilation calculation task windows

Marked events

ST events (lead changes, reference changes)

The following data is not transferred:

Waveforms, including ST and CO waveforms

Split Screen trends

Arrhythmia events

Data Transfer 10-5

Data Transfer

Page 301
Image 301
Philips V24CT, V26CT manual What is Transferred, Symbol Name Function Which Modules?