The Test Signals Function

Setting up your Monitor

System Self-Test Values:

Module

Test Numeric Limits

Test Waveforms

 

 

 

ECG and ECG/

 

 

RESP

 

 

ECG

100 bpm in ADULT mode

Simulated square wave and numeric

 

125 bpm in NEO/PEDI mode

 

RESP

15 rpm in ADULT mode

Simulated Resp wave and numeric

 

30 rpm in PEDI mode

 

 

55 rpm in NEO mode

 

 

 

 

Pressure

 

Simulated square wave and numerics

Systolic

120 mmHg ADULT

 

 

60 mmHg PEDI/NEO

 

Diastolic

0 mmHg ADULT/PEDI/NEO

 

 

 

 

SpO2/Pleth

 

 

SpO2

100%

Numeric only

Pleth

Wave on display

Simulated Wave

 

 

 

tcpO2/tcpCO2

 

Numeric only

tcpO2

60 mmHg (8.0 kPa)

 

tcpCO2

40 mmHg (5.3 kPa)

 

 

 

 

CO2

 

Simulated square wave and numerics

ETCO2

40 mmHg (6.0 kPa)

 

IMCO2

0 mmHg (0.0 kPa)

 

AWRR

25 rpm

 

 

 

 

Temperature

40°C (104°F)

Numeric only

 

 

 

3-60 Setting up your Monitor