Philips V24CT, V26CT manual System, Line protection fuses, T1.6 H Anesthetic Gas Exhaust

Models: V26CT V24CT

1 392
Download 392 pages 63.56 Kb
Page 347
Image 347

Controls and Connectors

Caution

The System Power Outlet cannot be used to output power to other devices such as the Philips M1167A/77A ACMS.

The System Power Outlet can only be used to supply power to the ACMS when the total leakage current does not exceed 300 µA.

4.Equipotential Grounding Terminal; this is used to connect the Anesthetic Gas Module to the hospital´s grounding system.

5.Line protection fuses, T1.6 H 250V.

6.Anesthetic Gas Exhaust.

C aution

Make sure that the anesthetic gas exhaust at the rear of the module is connected to the gas extraction system. If it is not, contact your biomedical engineer in order to make this connection.

The module can be operated from an ac power source of 100 - 240 V ± 10%, 50/60Hz.

The adjustment is made automatically by the power supply inside the module.

Assembling The Systems consist of the following separate parts:

the System

The Display Module (with control panel)

The Parameter Modules (with parameter racks)

The Computer Module

The Anesthetic Gas Module (optional)

If the system (display module and computer module) was delivered as an unpackaged shipment, the assembly procedure below does not apply.

Monitor Installation

and Patient Safety

 

 

Monitor Installation and Patient Safety 11-31

Page 347
Image 347
Philips V24CT, V26CT manual System, Line protection fuses, T1.6 H Anesthetic Gas Exhaust