Trends and Calculations

Viewing Patient Data

 

“Recording Functions” section (chapter 6) of this manual for further

 

information.

Selecting

You can select up to 9 parameters to be displayed in the Graph Trends

Parameters

Task Window by using the

 

 

softkey. The parameters

On / Off / Combine

d

for Graph

are placed into the graph channels according to the following rules:

Trends

 

 

 

 

If the current parameter is not selected and you press the On / Off / Combined key, the system will try to combine the parameter with any of the selected parameters in one graph channel. If this fails, a new graph channel will be assigned automatically.

If the current parameter is combined with any other parameter in a graph channel and you press the On / Off / Combined key, the system will try to assign this parameter to a new graph channel. If this fails, the parameter will be unselected.

If the current parameter is selected and is the only parameter in one graph channel, pressing the On / Off / Combined key will unselect this parameter from Graph Trends.

In the Vitals Signs Task Window, parameters that are selected for Graph Trends are displayed with a sign to the right of the parameter label:

Parameter is combined

A # sign appears next to the

 

parameter label

 

 

Parameter is on

A hyphen (--) appears next to the

 

parameter label

 

 

Parameter is off

No sign next to parameter label

 

 