Monitor Installation

and Patient Safety

 

 

Controls and Connectors

 

loudspeaker is not covered and can be heard. The most suitable

 

location for the External Alarm Device is in close proximity to the

 

ITE display.

The Rear

The connections on the rear panel of the Anesthetic Gas Module are

Panel of the

shown in the following diagram:

M1026A

 

Anesthetic

 

Gas Module

 

MONITOR

The rear of the Anesthetic Gas Module has the following connections:

1.Local Power connector; a 3-pin connector is used to input the local line voltage.

2.RS232 Connector (RS232 Interface); a 25-pin “D” type connector is used for connection to the ACMS.

Note—Do not apply a voltage greater than ±12V to the RS232 connections.

3.System Power Outlet (restricted use); this can be used to output power to the Philips M1165A/66A/75A/76A ACMS.

11-30 Monitor Installation and Patient Safety

Page 346
Image 346
Philips V26CT, V24CT manual Rear, M1026A Anesthetic Gas Module, ITE display, Shown in the following diagram