Troubleshooting

Troubleshooting

C aution

In a centrally networked system, if the previous patient is discharged only at the monitor and a new patient's data is transferred to the monitor, the new patient's name and ID will be overwritten by the central station's broadcast of the previous patient's name. To prevent this situation from occurring, the previous patient should be discharged from central station before the transfer.

The following table outlines the Data Transfer behavior under various events.

Event

Results

 

 

Power Off < 1 min

Transfer will restart automatically

 

 

Power Off > 1 min

Transfer is stopped. Collection of new data

 

must be restarted manually.

 

 

Op. Mode Change

Transfer is stopped. Data in module remains

 

intact.

 

 

Config Mode Change

Transfer is stopped. Data in module remains

 

intact.

 

 

Patient Discharge or End

Transfer is stopped. Data in monitor is

Case

deleted. Data in module remains intact.

 

 

Data Transfer 10-19