Performance Assurance Checks

Functional The system has a module self-test which generates and displays test

Testing waveforms and corresponding numerics for all the modules that are

Procedures connected and turned on. The test signals are displayed for about 30 seconds and then the display returns to monitoring mode. The test signal will be displayed only if the system self-test was successful and the parameter is plugged into the system and turned on.

Perform the following quick system check to verify that System board, integral rack and modules are communication properly before verifying the individual parameter modules.

Perform a You can verify that the System board, integral rack and modules are Quick System communicating properly by completing the following test.

Check

TEST - Press the ECG key on the front of the ECG module in a system that is turned On.

RESULT - Pressing the ECG key brings up the ECG Task Window and verifies that the System board and integral rack are communicating with the plug-in module.

If the plug-in module Task Window, waveform or numeric is missing from the display, the module may not be turned On. Try pushing the key on the plug-in to see if the system is communicating with the module.

If the ECG module does not respond, perform the module self-test.

To turn a parameter On, press the following key sequence:

1.Press Module Setup

2.Parametr On/Off

3.Select Parametr

4.On/Off Setup

Maintenance

Maintenance 13-13