Philips V26CT, V24CT manual Maintenance

Models: V26CT V24CT

1 392
Download 392 pages 63.56 Kb
Page 378
Image 378

Maintenance

Performance Assurance Checks

4.If you are using the V24CT or V26CT, make sure the battery Charging and Charged LEDs are behaving consistent with the battery charged. The following table describes the LED behavior:

LED Pattern During Charge Sequence

Residual Capacity

Charging LED

Charged LED

 

 

 

Up to 40%

Flashing

Off

 

 

 

More than 40%, Less than 90%

Steady

Off

 

 

 

More than 90%

Off

Steady

 

 

 

Full

Off

Steady

 

 

 

If the pattern is not correct, refer to the troubleshooting chapter of the Service Guide.

5.If you are using the V24CT/V26CT:

a.Disconnect the AC power cord. The V24CT or V26CT will switch to battery power and the display will indicate that a battery measurement is in progress. After about 30 seconds, a battery gauge and two battery symbols appear in the lower right corner of the screen display.

b.Remove one of the batteries. The monitor should continue to operate. If the monitor does not continue to operate go to “Battery Problems” in the Philips M1205A V24 and V26 Service Guide and check the fuses on the battery contact board.

c.Repeat this procedure for the second battery.

13-12 Maintenance

Page 378
Image 378
Philips V26CT, V24CT manual Maintenance