Using an ITE Display

EMC

1. The display must either fulfill the requirements of the EN60601-1-2

 

 

{IEC 601-1-2}, or

 

 

 

 

it must conform to the requirements of the EMC standard for ITE

 

 

devices, EN50082-1: 1997 and EN50081-1 / CISPR 22.

 

 

2. The video cable between M1167A/77A CMS Patient Monitoring

 

 

System and the ITE display must not exceed a length of 3.0m. For

 

 

cables exceeding 3.0m EMC testing according to IEC 801-4:1988 has

 

 

to be done.

 

 

Perfor-

Philips cannot assure compliance with the ANSI/AAMI EC-13 Standard for

mance

Cardiac Monitors, Heart-rate Meters and Alarms when using ITE displays.

Require-

Compliance with the ECG aspect ration and 25 mm/s ±10% sweep-speed

ments

specifications can only be assured when using the Philips Displays

 

 

M1092A, M1094B and M1095A. In addition, the following display

 

 

specifications are recommended:

 

 

 

 

 

 

 

Specification

 

Requirement or Value

Units

 

 

 

 

 

 

 

 

 

Resolution (pixel addressability)

1024 x 768

Dots x lines

 

 

 

 

 

Vertical Refresh Rate

60

Hz (non-interlaced)

 

 

 

 

 

Red, Green and Blue Video Inputs

~0.7

V p-p

 

 

 

 

 

Vertical & Horizontal Multi-Sync Inputs

5

V TTL

 

 

 

 

 

Video-Cable Connector

HD15male

 

 

 

 

 

 

The CMS and V24 and V26 Patient Monitors 1-39