Philips V24CT Cleaning the Roller on the Four Channel M1117A Recorder, Step, Equipment, Required

Models: V26CT V24CT

1 392
Download 392 pages 63.56 Kb
Page 227
Image 227

Loading Paper

Cleaning the Roller on the Four Channel (M1117A) Recorder

Step 3

Close the door. The paper should feed

 

automatically.

Note—The first sheet will come out folded, as shown.

If the paper doesn’t move through the roller properly, a possible cause is that the roller needs to be cleaned. (Be sure your hands are clean before touching the roller).

Equipment

Lint-freecloths or wipes

Required

Alcohol solvent that is low in water content. We recommend either

 

 

 

reagent grade denaturated ethanol or isopropanol (90% rather than

 

 

70%)

Directions 1. Turn the power off.

2.Open the door and remove any paper from the recorder.

3.Wipe the visible part of the roller with a saturated (but not dripping) cloth. Rotate the roller with clean fingertips to expose the entire surface and wipe the other side.

4.Remove any excess liquid by wiping with a dry cloth.

C aution

The roller must be completely dry before using the recorder or you can permanently damage the recording mechanism.

Recording Functions 6-55

Page 227
Image 227
Philips V24CT, V26CT manual Cleaning the Roller on the Four Channel M1117A Recorder, Step, Equipment, Required