Monitor Installation

and Patient Safety

 

 

Controls and Connectors

M1167/77A The display module (M1095A) cannot be attached to the computer System module (M1046B). It is designed for mounting on a horizontal or vertical

rail such as an I.V. pole. For further details, please refer to the CMS Service Guide, chapter 2; Installing CMS.

The computer module can be attached to the top or bottom of the Anesthetic Gas Module by a locking device. This is a lever which is operated from the rear of the monitor, and is attached to the bottom of both pieces of equipment.

To assemble the computer module and Anesthetic Gas Module:

Ensure that the locking lever is in the far right position.

Place the Anesthetic Gas Module onto the computer module, ensuring that its four feet are in their correct grooves.

Push the locking lever from right to far left. The equipment is now secure.

To dismantle the computer module and Anesthetic Gas Module:

Push the locking lever from left to far right.

Lift the Anesthetic Gas Module off the computer module.

M1165/66/ The display module (M1094B/92A) can be attached to the computer 75/76A module (M1046A) by a locking device. This is a lever which is operated System from the rear of the monitor, and is attached to the bottom of the display

module.

To assemble the computer module and display module:

Ensure that the locking lever is in the far right position.

Place the display module onto the computer module, ensuring that its four feet are in their correct grooves.

Push the locking lever from right to far left. The equipment is now secure.

To dismantle the computer module and display module:

Push the locking lever from left to far right.

Lift the display module off the computer module.

11-32 Monitor Installation and Patient Safety