11... APPENDIX J: TROUBLESHOOTING TIPS
of the Relay switches in the TEST position while observing the voltage. New batteries will read approximately 9 VDC and batteries below apx 7 volts are essentially dead.
OR... apply an external supply to the EXT power terminals on the
The OMP-MODL may have a corrupted Program Net:
Perform a system intialization with the following procedure:
A.Depress and hold the NEXT button down on the front panel.
B.Momentarily, depress the STOP and RESET buttons simultaneously, then release them.
C.After a second or so, release the NEXT button.
NOTE: This 3 button reset will erase all internal data as well as the currently loaded Program Net. Use with caution.
Power Failure / Batteries drain with External Power Source Connected:
The external power fuse may be blown. This would prevent the logger from running from the external power source and it would continue running from the batteries until they discharged. Refer to Appendix D for the location of this fuse. The fuse can be checked with a continuity meter and should read less than 1 ohm of resistance. If it is open, replace it with a 1.5A, Littelfuse P/N: 273.015 available from Omega Engineering Incorporated or an electronic components distributor.
SERIAL COMMUNICATION PROBLEMS:
Cannot establish an
Insure that only Omega Engineering
Check the Port number and that the port is actually working with another serial package and device (eg a modem).
Refer to addtional considerations in the HyperComm Serial
Communications chapter.
Modem Communication Problems:
Refer to Appendix K for modem configuration details.
Refer to the HyperComm Serial Communications chapter.
HYPERPLOT :
No Data Displays upon loading of File:
Data may actually be loaded but compressed. Select the
Calculations\Zoom All menu choice to view.
ANALOG READINGS:
Incorrect readings on
Ensure that the fuse for the particular channel is not blown.