3... INTERFACE MODULES

 

 

 

 

A

 

 

 

 

 

C

Excitation leads

1

2

3

4

5

6

7

8

9

10 11 12

 

RTD Element

 

 

 

 

 

 

 

 

 

 

 

+

 

 

 

 

 

 

 

 

 

 

+

 

 

 

 

 

 

 

 

 

 

-

 

 

 

 

 

 

 

 

 

 

-

 

 

 

 

 

 

Jumper

 

Sense leads

 

 

 

 

 

Cable Shield

 

 

 

 

 

 

Connect Shield to an Earth Ground

 

 

 

 

 

 

 

 

 

 

ML127

Figure 3... -30:

4-Wire Configuration

 

 

For long lead wire runs and in applications in electrically noisy environments, it is recommended that twisted pair and/or shielded wire be used. If shielded wire is used, the shield at the OMP-MODL end should be connected to an external earth ground (Figure 3... -

30)or if available, a grounded Shield connection provided on another type installed interface module (such as the MLIM-1).

Range

Excitation

Range

Excitation

 

Current

 

Current

 

 

 

 

Res 200 ohm

1 mA

Res 200,000 ohm

10 uA

Res 200 ohm

10 mA

Res 400,000 ohm

10 uA

Res 400 ohm

1 mA

RTD-100 ohm 300C

1 mA

Res 400 ohm

10 mA

RTD-100 ohm 850C

1 mA

Res 2000ohm

100 uA

RTD-1000 ohm 300C

100 uA

Res 4000 ohm

100 uA

RTD-1000 ohm 850C

100 uA

Res 10,000 ohm

100 uA

Therm -32 to 250C

10 uA

Res 20,000 ohm

100 uA

Therm -4 to 250C

10 uA

Res 40,000 ohm

10 uA

Therm +10 to 250C

10 uA

Res 100,000 ohm

10 uA

Therm +25 to 250C

100 uA

 

 

 

 

Excitation Currents used for MLIM-4 Ranges

USING THE OMP-MODL