Using Spectrum Analyzer Features

Saving and Recalling Data from Analyzer Memory

To Recall a Trace

1.Press (RECALL). If CARD is underlined, press INTERNAL CARD to select

INTERNAL.

2.Press Internal + Trace . This accesses a menu displaying TRACE A , TRACE B , TRACE C , LIMIT LINES, and AMP CDR.

3.Press TRACE A , TRACE B , or TRACE C to select the trace in which you want to place the trace data.

4.Enter the register number under which the trace was stored.

5.Press CENTER). The recalled trace is placed in the view mode and the cable TV analyzer state is changed to the state that was saved.

Trace data can also be recalled by using the catalog:

1.Press (RECALL). If CARD is underlined, press INTERNAL CARD to select

INTERNAL.

2 . P r e s s C a t a l o g I n t e r n a l , CATALOG REGISTER. Use the knob to highlight the number of the trace register to be retrieved. The trace registers have a “TR” preceding the trace register number.

3.Press LOAD FILE . The recalled trace is placed into trace B and the cable TV analyzer state is changed to the state that was saved.

3-11