Spectrum Analyzer Mode Key Descriptions

Spectrum Analyzer Mode Functions

LMT TEST ON OFF

LOAD

FILE

turns the limit-line testing and (if LMT DISP AUTO is selected) turns the display of the limit lines on and off. When limit-line testing is enabled, every measurement sweep of trace A is compared to the limit lines. If trace A is at or within the bounds of the limit lines, LIMIT PASS is displayed. If trace A is out of the limit-line boundaries, LIMIT FAIL is displayed.

Front-Panel Key Access cm)

loads a file from the memory card into analyzer memory. When the memory card is selected, pressing any of the catalog softkeys (CATALOG ALL , CATALOG STATES ,

CATALOG TRACES, CATALOG PREFIX, CATALOG DLP,

CATALOG AMP COR , CATALOG LMT LINE), or

CATALOG DISPLAY accesses LOAD FILE . When cataloging

analyzer memory using CATALOG REGISTER , press

LOAD FILE to recall the contents of a state or trace register into analyzer memory. To use the LOAD FILE function,

use the step keys to view sections of the directory, use the knob to select a file then press LOAD FILE . Trace data is loaded into trace B. See the softkey descriptions for CATALOG CARD and CATALOG REGISTER.

N O T E

Use of the LOAD FILE softkey is not recommended for recalling limit-line tables or amplitude-correction factors stored in analyzer memory

Front-Panel Key Access [RECALL) or (SAVE_)