Spectrum Analyzer Mode Key Descriptions

Spectrum Analyzer Mode Functions

SELECT TIME

allows you to enter the time value for a limit-line segment. The time value is with respect to the analyzer sweep time. A time value of zero is the start of the sweep, which is the left edge of the graticule. Enter the time value by using the data keys. Change the time value by using the step keys or the knob. Press [BKSPI) to correct errors.

N O T E

Limit-line data is sorted in time order in the limit-line table. The sorting occurs after you have entered the time and at least one amplitude value.

SELECT

TYPE

SELECT UPR AMPL

Front-Panel Key Access (j]

accesses the softkey menu used to select the limit-line type of line. Press FLAT to select a flat line, press SLOPE to

select a sloped line, or press POINT to select a point.

Front-Panel Key Access (j)

allows you to enter the amplitude value for the upper limit-line segment. Enter the amplitude value for the selected frequency or time by using the knob or data keys. Press [BK) to correct errors.

Front-Panel Key Access (j)