Using Spectrum Analyzer Features
Using Limit4ine Functions
Pressing EDIT DONE blanks the limit-line table from the screen and accesses the menu with LMT TEST ON OFF and LMT DISP Y N AUTO softkeys.
Pressing LMT TEST ON OFF turns the limit-line testing on and off.
Pressing LMT DISP Y N AUTO (Y) turns the limit lines display on. Pressing LMT DISP Y N AUTO (N) turns the limit lines display off. Pressing
LMT DISP Y N AUTO (AUTO) sets the limit line display to match the limit line test function. With AUTO underlined the limit lines are only displayed when limit line testing is turned on.
Pressing c-1 then Limit Lines accesses SAVE LIMIT and RECALL LIMIT . These softkey functions provide an easy way to save or
recall current limit-line tables. SAVE LIMIT saves the current limit-line tables in the current mass storage device (either cable TV analyzer memory or a memory card). To determine the current mass storage device, press SAVE LIMIT . If MAX REG # appears on the cable TV analyzer display, the current mass storage device is analyzer memory. If PREFIX= is displayed, the memory card is the mass storage device. Enter a register number, then press (ENTER] to save the current limit-line table in the current mass storage location.
RECALL LIMIT recalls limit-line tables from the current mass storage device (either cable TV analyzer memory or a memory card). To determine the current mass storage device, press RECALL LIMIT If MAX REG # appears on the cable TV analyzer display, the current mass storage device is analyzer memory. If PREFIX= is displayed, the memory card is the mass storage device. To recall a limit line, enter the register number that the limit-line table was saved under, then press [ENTER]. When recalling a limit line from the memory card, it is necessary that the current prefix matches the prefix that the limit line was stored with. Use Change Prefix to change the current prelix.