Using Spectrum Analyzer Features

Saving and Recalling Data from the Memory Card

To Recall a Trace

Press ISAVE) or [RECALL]. If INTERNAL is underlined, press INTER??AL CARD to select CARD.

Press Catalag Card then CATALOG TRACES Use the knob to highlight the trace data to be retrieved.

Press LOAD FILE . The trace data is placed in trace B.

Trace data can also be recalled by specifying the preEx and the register number :

1.Use Change Prefix to enter the preEx, or use the existing prefix.

2.Press (RECALL). If INTERNAL is underlined, press INTERNAL CARD to select CARD.

3.Press Card + Trace to access the menu that displays TRACE A , TRACE 3 , and TRACE C .

4.Select the trace in which you want the trace data stored by pressing TRACE A , TRACE B , or TRACE C

5.Enter the register number that the trace was saved under and then press (ENTER). The recalled trace is placed in view mode.

3-23